2014-06-04 00:41:00 | ALL | 999 | ALL | deployed by R/V Connecticut |
2014-08-31 02:00:00 | current_speed,current_u,current_v | 2 | aanderaa | Surface currents flagged as invalid due to suspected instrument problems (currents too low). Temperature and dissolved oxygen okay. |
2014-10-22 02:00:00 | dissolved_oxygen | 50 | sbe16 | Dissolved oxgyen data from 50m SBE16+ becoming erratic; flagged as invalid. |
2014-10-25 18:35:00 | ALL | 999 | ALL | Recovered by R/V Connecticut. 20m SBE37IM conductivity guard ripped off; 50m SBE16+ damaged (pump and O2 sensor dangling). |